"Products for the Generation, Detection and Measurement of Light"  

 

Home Up Notice Board Products Support Suppliers. Surplus Stock Site Map

 

SR-800

Extended Area Blackbody
The New ControlMaster SR-800 extended area blackbody from CI Systems produces superior accuracy and uniformity for reliable NIST traceable test and calibration results. As a result of its state-of-the-art digital control, the radiation heads may be easily interchanged for a more flexible system. The SR-800 operates as a complete standalone system or may be integrated into a larger test system.

   Superior accuracy for FLIR testing
   High uniformity for focal plane array (FPA) calibration
   Wide range of temperatures
   Interchangeable heads with the same controller
   Radiation heads independent of the controller for easy maintenance     and calibration
   Robust built-in tests to shorten maintenance time
   Calibration procedure integrated into the controller
   Certified to MIL-T-28800D and CE
   Compact portable controller

 

Specifications

SR-800 Model

2D/A

4D/A

7D/A

12 D/A

14 D/A

Aperture2”
diameter
4”x4”7”x7”12”x12”14”x14”

Differential Temp. Range

-20ºC
to
75ºC

-25ºC
to
+100ºC

-20ºC
to
+65ºC

-15ºC
to
+55ºC

-15ºC
to
+55ºC

Absolute
Temp. Range

5ºC
to
100ºC

0ºC
to
+125ºC

5ºC
to
+90ºC

10ºC
to
+80ºC

10ºC
to
+80ºC

 
Specifications

SR-800 Model

2D/A

4D/A

7D/A

12 D/A

14 D/A

Differential Temp. Accuracy
@ |ΔT| < 10ºC

± 0.003ºC

± 0.003ºC

± 0.003ºC

± 0.003ºC

± 0.003ºC

Absolute Temp. Accuracy

± 0.025ºC

± 0.025ºC

± 0.025ºC

± 0.025ºC

± 0.025ºC

Long Term Stability @ 0°C & 50°C Ambient

± 0.001ºC

± 0.001ºC

± 0.001ºC

± 0.002ºC

± 0.002ºC

Short Term Stability @|ΔT| < 10ºC

± 0.002ºC

± 0.002ºC

± 0.002ºC

± 0.002ºC

± 0.002ºC

Uniformity @ |ΔT| < ±5ºC

0.01ºC

0.01ºC

0.03ºC

0.06ºC

0.06ºC

Emissivity

0.97

0.97

0.97

0.97

0.97

Settling Time to ±0.01°C of set-point  at ±10ºC step

60 seconds

60 seconds

90 seconds

90 seconds

90 seconds

Options

SR-800 Model

2D/A

4D/A

7D/A

12 D/A

14 D/A

[ET] Extended Differential Temp.

-----

-30ºC to+150ºC

-20ºC to +150ºC

-20ºC to +100ºC

-----

[ET] Extended Absolute Temp.

-----

-5ºC to +175ºC

5ºC to +175ºC

5ºC to +125ºC

-----

[HU] High uniformity
@ |ΔT| < ±5ºC

-----

-----

0.01ºC

0.028ºC

-----

[WTR] Absolute Temp. Range

-----

-20ºC to 45ºC

-----

-----

-----

[CH] Differential Temp. Range

-10ºC to 40ºC

-10ºC to 40ºC

-10ºC to 40ºC

-----

-----

[CH] Absolute Temp. Range

-40ºC to 80ºC

-40ºC to 80ºC

-40ºC to 80ºC

-----

-----

Options Legend
[WTR] - Wide
Temperature Range
[HMIS] - Extended High
Emissivity (0.99)
[CH] - In Climate Chamber
Operation [-30°C to 80°C]
[GPIB] - GPIB Communication
[TH] - Two radiation heads
operated simultaneously by
the same controller
[USB] - USB Communication
[TW] - Up to four
motorized target wheels
[RM] - 19" Rack mount kit
General Specifications
Readout resolution0.001 ºC
Operating voltage90 - 264 VAC, 50/60 Hz
Operating temperature0ºC - 50ºC
Controller dimensions342x310x133 mm